    Design criteria of a transcutaneous power delivery system for implantable devices.

    Get PDF
    Implantable cardiac assist devices such as artificial hearts and blood pumps are a rapidly growing therapy used for treating moderate to severe congestive heart failure. While current treatments offer improved heart failure survival and increased patient functionality with enhanced quality of life, powering these devices are still constraining. In practice, percutaneous cables passing through skin are used for power and control data transmission requiring patients to maintain a sterile dressing on the skin cable-exit site. This contact site limits patient movement as it is vulnerable to wound infection due to trauma and poor healing. As a result, a sterile dressing has to be maintained and nursed regularly for treating the wound. Complications from the exit site infections are a leading cause of death in long-term support with these devices. Wireless power and control transmission systems have been studied and developed over years in order to avoid percutaneous cables while supplying power efficiently to the implanted device. These power systems, commonly named Transcutaneous Energy Transfer (TET) systems, enable power transmission across the skin without direct electrical connectivity to the power source. TET systems use time-varying electromagnetic induction produced by a primary coil that is usually placed near skin outside the body. The induced voltage in an implanted secondary coil is then rectified and regulated to transfer energy to an implanted rechargeable battery in order to power the biomedical load device. Efficient and optimum energy transfer using such transcutaneous methods is more complex for mobile patients due to coupling discrepancies caused by variations in the alignment of the coil. The research studies equivalent maximum power transfer topologies for evaluating voltage gain and coupling link efficiency of TET system. Also, this research adds to previous efforts by generalizing different scenarios of misalignments of different coil size that affects the coupling link. As a whole, this study of geometric coil misalignments reconsiders potential anatomic location for coil placement to optimize TET systems performance in anticipated environment for efficient and safe operation.--Abstract

    Enhancement of Solar PV Hosting Capacity in a Remote Industrial Microgrid: A Methodical Techno-Economic Approach

    Get PDF
    To meet the zero-carbon electricity generation target as part of the sustainable development goals (SDG7), remote industrial microgrids worldwide are considering the uptake of more and more renewable energy resources, especially solar PV systems. Estimating the grid PV hosting capacity plays an essential role in designing and planning such microgrids. PV hosting capacity assessment determines the maximum PV capacity suitable for the grid and the appropriate electrical location for PV placement. This research reveals that conventional static criteria to assess the PV hosting capacity fail to ensure the grid’s operational robustness. It hence demands a reduction in the theoretical hosting capacity estimation to ensure grid compatible post-fault voltage and frequency recovery. Energy storage technologies, particularly fast-responsive batteries, can potentially prevent such undesirable scenarios; nevertheless, careful integration is required to ensure an affordable cost of energy. This study proposes a novel methodical techno-economic approach for an off-grid remote industrial microgrid to enhance the PV hosting capacity by integrating battery energy storage considering grid disturbance and recovery scenarios. The method has been validated in an industrial microgrid with a 2.6 MW peak demand in a ready-made garment (RMG) factory having a distinctive demand pattern and unique constraints in remote Bangladesh. According to the analysis, integrating 2.5 MW of PV capacity and a 1.2 MVA battery bank to offset existing diesel and grid consumption would result in an energy cost of BDT 14.60 per kWh (USD 0.1719 per kWh). For high PV penetration scenarios, the application of this method offers higher system robustness, and the financial analysis indicates that the industries would not only benefit from positive environmental impact but also make an economic profit
    corecore